ECG trigger monitor & cables

Cardiac Trigger Monitor

ECG trigger for use in synchronized application in Nuclear and Molecular Imaging such as Gated-SPECT and MUGA studies.

  • intuitive operation and reliable cardiac trigger
  • user specific pulse amplitude options of 1, 50, 100 or 150 ms
  • active Matrix TFT 6.5“ (16.5 cm) Color Touch Screen with a 640 x 480 Resolution
Technical Specifications:
  • Voltage Input: 100 -120 V~ / 200-230 V~
  • Line Frequency: 50 / 60 Hz
  • Max. Power Consumption: 45 VA
Dimensions:
  • Height: 7.49“ (19.02 cm)
  • Width: 7.94“ (20.17 cm)
  • Depth: 5.18“ (13.16 cm)
  • Weight: 3.9 lbs (1.8 kg)
Technical Specifications:
Patient Cable

4 Lead cable with 6 PIN AAMI standard connector, isolated from ground related cir- cuits by > 4 kV rms, 5.5 kV peak

Frequency Response

LCD Display and Recorder

Filtered 1.5 to 40 Hz or 3.0 to 25 Hz

Unfiltered 0.67 to 100 Hz

ECG Output Unfiltered 0.67 to 100 Hz

Electrode Impedance Measurement

Technique 10 Hz AC-Signal < 10 μA rms

Range Range 200 kΩ per lead

Accuracy ±3 % ±1 kΩ

Recommended Electrode

10 % Chloride sponge type

Synchronized Output (Trigger)*

Output Trigger Delay < 2 ms

R to R Trigger Accuracy ±75 µs @ 1 mV input

Pulse Width Options 1 ms, 50 ms, 100 ms or 150 ms

Pulse Amplitude Options 0 V to +5 V or -10 V to +10 V

Output Impedance < 100 Ω

*Test input signal at ECG leads Conditions: 1/2 sine wave, 60 ms width, 1 mV amplitude, 1 pulse/second

Alarms

High Rate 15 – 250 bpm (5 bpm increments)

Low Rate 10 – 245 bpm (5 bpm increments)