TECHNICAL DATA
Detector Assembly |
Detector-Material |
Lutetium Oxyorthosilicate (LSO) |
Crystal Dimensions |
6.45 x 6.45 x 25 mm |
|
Crystals per Detector Block |
64 |
|
Detector Ring Diameter |
824 mm (32.4 in.) |
|
Total Number of Detectors |
9,216 |
|
Axial Field-of-view |
162 mm (6.38 in.) |
|
Transversal Field of view |
583 mm (23 in.) |
|
Performance Specifications |
Spatial Resolution |
5.7 mm |
Sensitivity @ 425 keV |
4.9 cps / kBq |
|
Non Uniformity |
< 5 % variation |
|
Count Rate Peak NECR |
100 Kcps @ 33 kBq / cc |
|
Coinsidence Time Window |
6 nsec |
|
Gantry Specifications |
Height |
188 cm (74 in.) |
Width |
215 cm (72 in.) |
|
Depth (without PHS) |
81 cm (32 in.) |
|
Weight |
1270 kg (2806 lb.) |
|
Patient Port Diameter |
562 mm (22.1 in.) |
|
2D Transmission |
3 retractable, self storing Ge68 sources |
|
Patient Handling Sytem (PHS) |
Height |
57-93 cm (22-36 in.) |
Width |
51 cm (20 in.) |
|
Depth |
281 cm (110 in.) |
|
Weight |
386 kg (850 lb.) |
|
Maximum patient weight |
227 kg (500 lb.) |
Acquisitian & Processing Hardware
New – Reliable – Available
PowerPC for Acquisition and high-end system interfaces that replaces the ACS and SUN, which allows for additional storage capacity, while decreasing scan and reconstruction times. All SCINTRON components are new and industry parts for long-term availability.
- New Hard- and Software Technology
- Parallel Acquisition and Processing
- Faster Acquisition and Reconstruction
- List-Mode Acquisitions
- 3D Cardiac Acquisitions
- Reduced Patient Dose
- DICOM Interface
- Longevity
- Remote Support